138
Fluke patient simulator modification
Welch Allyn Propaq CS Vital Signs Monitor
Incompatibilities with other simulators may also cause erroneous readings and/or missing 
ECG. Check Propaq CS monitor readings with the simulator against a known source 
before using the simulator for calibration.
Function
Color
Pin No.
 Propaq CS 244/246
Output (+)
Black
4
C
Output (-)
Red
1
A & E
Exciter (+)
White
3
B
Exciter (-)
Green
5
D
ECG Ref
Blue
2
No Connection
810-0696-57
W1
W2