Programming the VT1422A for Data Acquisition and Control   133
The Operating
Sequence
The VT1422A has four major operating phases plus one optional phase. 
Figure 4-7 shows these phases. A trigger event starts the sequence:
1. (INPUT): the state of all digital inputs are captured and each analog 
input channel that is in the scan list and/or referenced by an algorithm 
variable is scanned. Reading values from channels placed in the Scan 
List with ROUT:SEQ:DEF are sent to the CVT and/or FIFO.
1A.(Remote Runtime Scan Verification): If a Scan Status Variable (S1xx) 
is referenced in any algorithm, this time is used to check the scan list 
execution of the Remote Signal Conditioning Unit (RSCU) connected 
to the channel xx. The S1xx variable will then take on one of three 
values; 0=normal operation, 1=the RSCU is disconnected and 3= the 
RSCU scan list was out of synchronization. Each VT1539A SCP has 
2 main channels so there are 16 possible scan status variables; S100, 
S101, S108, S109, S116, S117, S124, S125, S132, S133, S140, S141, 
S148, S149, S156, and S157. If no S1xx variable is referenced in any 
algorithm, then phase 1A is not executed.
2. (UPDATE): The update phase is a window of time made large enough 
to process all variables and algorithm changes made after INIT. Its 
width is specified by ALG:UPDATE:WINDOW. This window is the 
only time variables and algorithms can be changed. Variable and 
algorithm changes can actually be accepted during other phases, but 
the changes don’t take place until an ALG:UPDATE command is 
received and the update phase begins. If no ALG:UPDATE command 
is pending, the update phase is simply used to accept variable and 
algorithm changes from the application program (using ALG:SCAL, 
ALG:ARR, ALG:DEF). Data acquired by external specialized 
measurement instruments can be sent to the algorithms at this time.
Note
Changing algorithm variables requires VT1422A hardware resources that 
can only be provided during the INPUT and UPDATE phases of the 
operating cycle. The VT1422A does not update variables during the time 
between the CALCULATE and OUTPUT phases. Therefore, applications 
that are intensive in the update area should consciously extend the INPUT 
and UPDATE periods through use of the ALG:UPD:WINDOW and 
SAMP:TIME commands or by reducing the time between the CALCULATE 
and OUTPUT phases through shorter algorithm loop time. See TRIG:TIMer.
3. (CALCULATE): all INPUT and UPDATE values have been made 
available to the algorithm variables and each enabled algorithm is 
executed. The results to be output from algorithms are stored in the 
Output Channel Buffer.
4. (OUTPUT): each Output Channel Buffer value stored during 
(CALCULATE) is sent to its assigned SCP channel. The start of the 
OUTPUT phase relative to the Scan Trigger can be set with the SCPI 
command ALG:OUTP:DELay.