144 Programming the VT1422A for Data Acquisition and Control  
Status Bit Descriptions
Questionable Data Group
Bit 
 Bit Value
 Event Name
 Description
8
 256
 Lost Calibration
At *RST or Power-on Control Processor has found a checksum error in the 
Calibration Constants. Read error(s) with SYST:ERR? query and re-calibrate areas 
that lost constants.
9
 512
 Trigger Too Fast
Scan not complete when another trigger event received.
10
 1024
 FIFO Overflowed
Attempt to store more than 65,024 values in FIFO. 
11
 2048
 Overvoltage 
(Detected on Input)
If the input protection jumper has not been cut, the input relays have been opened 
and *RST is required to reset the module. Overvoltage will also generate an error.
12
 4096
 VME Memory 
Overflow
The number of values taken exceeds VME memory space.
13
 8192
 Setup Changed
Channel Calibration in doubt because SCP setup may have changed since last 
*CAL? or CAL:SETup command. (*RST always sets this bit.)
Operation Status Group
Bit 
 Bit Value
 Event Name
 Description
0
 1
 Calibrating
Set by CAL:TARE and CAL:SETup. Cleared by CAL:TARE? and CAL:SETup?. Set 
while *CAL? executing, then cleared.
4
 16
 Measuring
Set when instrument INITiated. Cleared when instrument returns to Trigger Idle 
State.
8
 256
 Scan Complete
Set when each pass through a Scan List is completed.
9
 512
 SCP Trigger
Reserved for future SCPs.
10
 1024
 FIFO Half Full
FIFO contains at least 32,768 values.
11
 2048
 Algorithm Interrupt
The interrupt() function was called in an executing algorithm.
Standard Event Group
Bit 
 Bit Value
 Event Name
 Description
0
 1
 Operation Complete
*OPC command executed and instrument has completed all pending operations.
1
 2
 Request Control
Not used by VT1422A
2
 4
 Query Error
Attempting to read empty output queue or output data lost.
3
 8
 Device Dependent Error
A device-dependent error occurred. See Appendix B page 453.
4
 16
 Execution Error
Parameter out of range or instrument cannot execute a proper command because 
it would conflict with another instrument setting. 
5
 32
 Command Error
Unrecognized command or improper parameter count or type.
6
 64
 User Request
Not used by VT1422A.
7
 128
 Power-On
Power has been applied to the instrument.