158 Programming the VT1422A for Data Acquisition and Control  
4. If some improvement is seen, increase the sample period again and 
perform another test. When the sample period is increased and no 
improvement is seen, the maximum settling delay that any single 
channel requires has been found.
5. If the quality of the measurements does not respond to this increase in 
sample period, then inadequate settling time is not likely to be 
causing measurement problems.
Fixing the Problem
If the system scans fast enough with the increased sample period, the 
problem is solved. The system is only running as fast as the slowest channel 
allows, but, if its fast enough, that’s okay. If, on the other hand, getting 
quality readings has slowed the scan rate too much, there are two other 
methods that will, either separately or in combination, have the system 
making good measurements as fast as possible.
Use Amplifier SCPs
Amplifier SCPs can remove the need to increase settling delays. How? 
Each gain factor of four provided by the SCP amplifier allows the Range 
Amplifier to be set one range higher and still provide the same measurement 
resolution. Amplifier SCPs for the VT1422A are available with gains of 0.5, 
8, 16, 64, and 512. Now return to the earlier example of a difficult 
measurement where one channel is measuring 15.5 volts on the 16 volt 
range and the next a thermocouple on the 0.0625 range. If our thermocouple 
channel is amplified through an SCP with a gain of 16, the Range Amplifier 
can be set to the 1 volt range. On this range the A/D resolution drops to 
around 31 µV per LSB so the stray capacitances discharging after the 
15.5 volt measurement are now only one-sixteenth as significant and thus 
reduce any required settling delay. Of course for most thermocouple 
measurements, a gain of 64 can be used and set the Range Amplifier to the 
4 volt range. At this setting, the A/D resolution for one LSB drops to about 
122 µV and further reduces or removes any need for additional settling 
delay. This improvement is accomplished without any reduction of the 
overall measurement resolution.
 NOTE
Filter-amplifier SCPs can provide improvements in low-level signal 
measurements that go beyond just settling delay reduction. Amplifying the 
input signal at the SCP allows using less gain at the Range Amplifier (higher 
range) for the same measurement resolution. Since the Range Amplifier has 
to track signal level changes (from the multiplexer) at up to 100 kHz, 
its bandwidth must be much higher than the bandwidth of individual 
filter-amplifier SCP channels. Using higher SCP gain along with lower 
Range Amplifier gain can significantly increase normal-mode noise rejection.