VXi VT1422A 사용자 설명서
VT1422A Command Reference 245
ALGorithm
ALGorithm[:EXPLicit]:SCALar?
ALGorithm[:EXPLicit]:SCALar? ’<alg_name>’,’<var_name>’
returns the value of
the scalar variable <var_name> in algorithm <alg_name>.
Parameters
Comments
•
An error is generated if <alg_name> or <var_name> is not defined.
•
Returned Value: numeric value. The type is float32.
•
Use VXIplug&play function: hpe1422_algExpScal_Q(...)
ALGorithm[:EXPLicit]:SCAN:RATio
ALGorithm[:EXPLicit]:SCAN:RATio ’<alg_name>’,<num_trigs>
specifies the
number of scan triggers that must occur for each execution of algorithm <alg_name>.
This allows the specified algorithm to be executed less often than other algorithms.
This can be useful for algorithm tuning.
This allows the specified algorithm to be executed less often than other algorithms.
This can be useful for algorithm tuning.
Notes
1. The command ALG:SCAN:RATio <alg_name>,<num_trigs> does not take
effect until an ALG:UPDATE or ALG:UPD:CHAN command is received.
This allows multiple ALG:SCAN:RATIO commands to be sent with their
effect synchronized with an ALG:UPDATE command.
This allows multiple ALG:SCAN:RATIO commands to be sent with their
effect synchronized with an ALG:UPDATE command.
2. ALG:SCAN:RATio places a variable update request in the Update Queue.
More update requests cannot be placed in the Update Queue than are allowed
by the current setting of ALG:UPD:WINDOW or a "Too many updates ––
send ALG:UPDATE command" error message will be generated.
by the current setting of ALG:UPD:WINDOW or a "Too many updates ––
send ALG:UPDATE command" error message will be generated.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
alg_name
string
ALG1 - ALG32
none
var_name
string
valid 'C' variable name
none
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
alg_name
string
ALG1 - ALG32
none
num_trigs
numeric (int16)
1 to 32,767