278 VT1422A Command Reference  
DIAGnostic
DIAGnostic:CALibration:TARE[:OTDetect]:MODE
DIAGnostic:CALibration:TARE[:OTDetect]:MODE <mode
sets whether Open 
Transducer Detect current will be turned off or left on (the default mode) during the 
CAL:TARE operation.
Parameters
Comments
When <mode> is set to 0 (the *RST Default), channels are tare calibrated with 
their OTD current off. When <mode> is 1, channels that have their OTD current 
on (DIAGnostic:OTDetect ON,(@<ch_list>)) are tare calibrated with their 
OTD current left on.
By default (*RST) the CALibration:TARE? command will calibrate all 
channels with the OTD circuitry disabled. This is done for two reasons. 
First, most users do not leave OTD enabled while taking readings; and second, 
the CALibration:TARE? operation takes much longer with OTD enabled. 
However, for users who intend to take readings with OTD enabled, setting 
DIAG:CAL:TARE:OTD:MODE to 1, will force the CAL:TARE? command to 
perform calibration with OTD enabled on channels so specified by the user with 
the DIAG:OTD command.
Related Commands: *CAL?, CAL:SET, DIAG:CAL:SET:MODE?
*RST Condition: DIAG:CAL:TARE:MODE 0
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
configure OTD on during CAL:TARE
DIAG:CAL:TARE:MODE 1
set mode for OTD to stay on
CAL:TARE?
start channel tare cal.
DIAGnostic:CALibration:TARE[:OTDetect]:MODE?
DIAGnostic:CALibration:TARE[:OTDetect]:MODE? 
 returns the currently set 
mode for controlling Open Transducer Detect current while performing CAL:TARE? 
operation.
Comments
Returns a 0 when OTD current will be turned off during CAL:TARE?. Returns 
1 when OTD current will be left on during CAL:TARE? operation. The data 
type is int16.
Related Commands: DIAG:CAL:TARE:MOD, DIAG:OTD, CAL:TARE?
*RST Condition: DIAG:CAL:TARE:MODE 0
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
mode
boolean (uint 16)
 0 | 1
volts