306 VT1422A Command Reference  
MEASure
MEASure:VOLTage:UNSTrained?
MEASure:VOLTage:UNSTrained?  (@<ch_list>) 
This command automatically 
configures the VT1422A to measure the bridge output voltage at each channel in 
<ch_list> and initiates a measurement scan. It averages 32 measurements for each 
channel and the averaged values are stored for later use by the strain Engineering 
Units conversion process on these channels. The strain bridges must be unstrained 
during this time. The average of each channel’s reading is also sent to the FIFO buffer 
to be viewed later if desired. The command returns a single value which is the number 
of readings sent to the FIFO.
Note that this command executes a measurement scan without executing any 
algorithms that might be defined. 
The sequence of individual commands to approximate this operation is
TRIGger:COUNt 1
one time through scan list
ROUTe:SEQuence:DEFine (@<ch_list>)
input the list of channels to measure
SENSe:FUNCtion:VOLTage [<range>,](@<ch_list>set measurement function to volts
SENSe:STRain:EXCitation:STATE ON,(@<ch_list>turn on excitation supplies
SENSe:STRain:CONNect BRIDge,(@<ch_list>)
connect channel sense to bridge output
INIT
start measurement scan
SENSe:DATA:FIFO:COUNT?
query for number of readings in FIFO
enter statement here to return FIFO reading <count>
SENSe:DATA:FIFO:PART? <count>
read unstrained voltage readings from 
the FIFO buffer
enter statement to for block data from FIFO
next the unstrained voltage values acquired above must be sent back to the VT1422A by 
executing the following command once for each channel in <ch_list> above:
SENSe:STRain:UNSTrained <voltage_value>,(@channel)
Notes
1. Unlike the MEAS:VOLT:UNST? command, the individual command 
sequence above cannot keep defined algorithms from running at INIT. 
Since algorithms can place values into the FIFO buffer, it will be necessary to 
determine which FIFO values are the excitation voltages. 
2. Remember that the MEAS:VOLT:UNST? command also provides the 
average of 32 measurements for each excitation value.
Parameters
Comments
This command is only for use on channels measured with the VT1529A/B. If 
executed on channels connected to other strain SCPs, a 3007 "Invalid signal 
conditioning plug-on" error message will be generated.
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
ch_list
channel list (string)
10000 - 15731
none