322 VT1422A Command Reference  
OUTPut
OUTPut:VOLTage:AMPLitude
OUTPut:VOLTage:AMPLitude <amplitude>,(@<ch_list>) 
sets the excitation 
voltage on programmable Strain Bridge Completion SCPs pointed to by <ch_list
(the VT1511A for example).
Note
This command is not used to set output voltage on SCPs like the VT1531A Voltage 
Output SCP.
Parameters
Comments
To turn off excitation voltage (when using external voltage source) program 
<amplitude> to 0.
Related Commands: OUTP:VOLT:AMPL?
*RST Condition: MIN (0) for VT1511A
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
OUTP:VOLT:AMPL 5,(@116:119)
set excitation voltage for channels 16 
through 19
OUTPut:VOLTage:AMPLitude?
OUTPut:VOLTage:AMPLitude? (@<channel>) 
returns the current setting of 
excitation voltage for the channel specified by <channel>. If the channel is not on a 
VT1511A SCP, the query will return zero.
Comments
The <channel> parameter must specify a single channel only.
Returned Value: Numeric, one of 0, 1, 2, 5, or 10 for VT1511A SCP, 3.9 for 
non- programmable VT1506A/07A SCPs. Data type is float32
Related Commands: OUTP:VOLT:AMPL
Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)
Usage
OUTP:VOLT:AMPL? (@103)
returns current setting of excitation 
voltage for channel 3
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
amplitude
numeric (float32)
MIN | 0 |1 | 2 | 5 | 10 | MAX
none
ch_list
channel list (string)
100 - 163
none