VXi VT1422A 사용자 설명서

다운로드
페이지 529
328 VT1422A Command Reference  
SAMPle
 SAMPle
The SAMPle subsystem provides commands to set and query the interval between 
channel measurements (pacing).
Subsystem Syntax
SAMPle
:TIMer  <interval>
:TIMer?
SAMPle:TIMer
SAMPle:TIMer  <interval
sets the time interval between channel measurements. 
Parameters
Comments
The minimum <interval> is 40 µs. The resolution for <interval> is 2.5 µs.
If the Sample Timer interval multiplied by the number of channels in the 
specified Scan List is longer than the Trigger Timer interval, at run time a 
"Trigger too fast" error will be generated.
the SAMP:TIMER interval can change the effect of the 
SENS:CHAN:SETTLING command. SENS:CHAN:SETT specifies the 
number of times a channel measurement should be repeated for channels defined 
in an algorithm. The total settling time per channel then is (SAMP:TIMER 
<interval>) X (<chan_repeats> from SENS:CHAN:SETT)
When Accepted: Not while INITiated
Related Commands: SENSE:CHAN:SETTLING, SAMP:TIMER?
*RST Condition: Sample Timer for all Channel Lists set to 4.0E-5 seconds.
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SAMPLE:TIMER  150E-6
Pace measurements at 150 µs intervals 
channel to channel
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
interval
numeric (float32)
(string)
4.0E-5 to 16.3825E-3 |
MIN | MAX
seconds