352 VT1422A Command Reference  
[SENSe]
If using amplifier SCPs, set them first and keep their settings in mind when 
specifying a range setting. For instance, if the expected signal voltage is to be 
approximately 0.1 V dc and the amplifier SCP for that channel has a gain of 8, 
<range> must be set no lower than 1 V dc or an input out-of-range condition 
will exist.
The <sub_type> parameter: values of 85 and 92 differentiate between 100 
Ω (@ 
0
°C) RTDs with temperature coefficients of 0.00385 and 0.00392 ohm/ohm/°C 
respectively. The <sub_type> values of 2250, 5000, and 10000 refer to 
thermistors that match the Omega 44000 series temperature response curve. 
These 44000 series thermistors are selected to match the curve within 0.1 or 
0.2
°C. For thermistors, <sub_type> may be specified in kΩ (kohm).
The <sub_type> EEXTended applies to E type thermocouples at 800
°C and 
above.
CUSTom is pre-defined as Type K, with no reference junction compensation 
(reference junction assumed to be at 0
°C).
The *CAL? command calibrates temperature channels based on Current Source 
SCP and Sense Amplifier SCP setup at the time of execution. If SCP settings 
are changed, those channels are no longer calibrated. *CAL? must be executed 
again.
When Accepted: Not while INITiated
Related Commands: *CAL?, OUTP:CURR (for RTDs and Thermistors), 
SENS:REF and SENS:REF:TEMP (for Thermocouples) 
*RST Condition: SENSE:FUNC:VOLT AUTO on all channels and bridges set 
to FBEN on all strain channels
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
Link two channels to the K type thermocouple temperature conversion
SENS:FUNC:TEMP  TCOUPLE,K,(@101,102)
Link channel 0 to measure reference temperature using 5k thermistor
SENS:REF  THER,5000,(@100)