VXi VT1422A 사용자 설명서

다운로드
페이지 529
366 VT1422A Command Reference  
[SENSe]
[SENSe:]STRain:EXCitation:STATe
[SENSe:]STRain:EXCitation:STATe <enable>,(@<ch_list>) 
connects or 
disconnects all four excitation supply ports on a VT1529A/B. Only one channel for 
each VT1529A/B needs to be specified in <ch_list> and all four excitation supply 
ports on that unit will configure as specified in <enable>. The first channel number 
on each possible VT1529A/B is: 10000, 10100, 10800, 10900, 11600, 11700, 12400, 
12500, 13200, 13300, 14000, 14100, 14800, 14900, 15600, 15700.
Parameters
Comments
Related Commands: SENSE:STRAIN:
…, SENSE:FUNC:STRAIN…
*RST Condition: OFF
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
STRAIN:EXC:STAT ON,(@10800)
turn on all excitation supplies on 
VT1529A/B connected to on-board 
channel 08
[SENSe:]STRain:EXCitation:STATe?
[SENSe:]STRain:EXCitation:STATe? (@<channel>) 
returns the state of all four 
VT1529A/B excitation supply ports referenced in <channel>. The first channel 
number on each possible VT1529A/B is: 10000, 10100, 10800, 10900, 11600, 11700, 
12400, 12500, 13200, 13300, 14000, 14100, 14800, 14900, 15600, 15700.
Parameters
Comments
Related Commands: SENSE:STRAIN:EXC:STAT
Returned Value: Numeric, 0 or 1. Type is uint16.
*RST Condition: OFF
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
Usage
STRAIN:EXC:STAT? (@10800)
check state of excitation supply ports on 
VT1529A/B connected to channel 08
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
enable
boolean (uint16)
ON | OFF
none
ch_list
channel list (string)
10000 - 15731
none
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
channel
channel list (string)
1000 - 15731
none