VT1422A Command Reference   395
TRIGger
TRIGger:COUNt
TRIGger:COUNt  <trig_count
sets the number of times the module can be 
triggered before it returns to the Trigger Idle State. The default count is 1. Note that 
this default was chosen to make testing data acquisition scan lists easier (only one 
scan list worth of data in FIFO per trigger). For algorithm operation, it is usually 
desirable to change the count to INFinite to accept continuous triggers. See Figure 7-7 
on page 394.
Parameters
Comments
When <trig_count> is set to 0 or INF, the trigger counter is disabled. Once 
INITiated the module will return to the Waiting-For-Trigger State after each 
trigger event. The ABORT (preferred) and *RST commands will return the 
module to the Trigger Idle State. ABORT is preferred since *RST also returns 
other module configurations to their default settings.
The default count is 0.
Related Commands: TRIG:COUNT?
*RST Condition: TRIG:COUNT 0
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
TRIG:COUNT  10
Set the module to make ten passes all 
enabled algorithms.
TRIG:COUNT  0
Set the module to accept unlimited 
triggers (the default)
TRIGger:COUNt?
TRIGger:COUNt? 
returns the currently set trigger count.
Comments
If TRIG:COUNt? returns 0, the trigger counter is disabled and the module will 
accept an unlimited number of trigger events.
Returned Value: Numeric 0 through 65,535. The data type is int32.
Related Commands: TRIG:COUNT
*RST Condition: TRIG:COUNT? returns 0
Send with VXIplug&play Function: hpe1422_cmdInt32_Q(...)
Usage
TRIG:COUNT?
Query for trigger count setting
enter statement
Returns the TRIG:COUNT setting
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
Trig_count
numeric (uint16)
(string)
0 to 65535 | INF
none