Programming the VT1422A & VT1529A/B for Remote Strain Measurement   87
6. Turn on excitation voltage to the strain bridges with the 
SENSe:STRain:EXCitation:STATe ON,(@<ch_list>) command.
7. Use the ROUTe:SEQuence:DEFine (@<ch_list>) command to 
define the scan list to measure the output voltage at each strain bridge. 
The <ch_list> specified here must match the <ch_list> specified in 
the two previous steps (measuring unstrained and excitation voltages).
8. Start the measurement scan with the INIT command. The default 
trigger system settings will execute a single measurement scan. 
During the scan, each channel reading is sent to the FIFO and CVT. 
The readings are then retrieved and used to calculate the strain for 
each channel using the excitation, unstrained, and strained voltage 
values.
Figure 3-18 shows the sequence of commands to convert bridge voltage 
measurements to strain by post-processing.