VXi VT1422A 사용자 설명서

다운로드
페이지 529
Programming the VT1422A & VT1529A/B for Remote Strain Measurement   93
Error Analysis
Figure 3-21 compares the non-linear quarter bridge equation used for strain 
SCPs with the linear approximation used with the VT1529A/B. Notice that 
while the error is independent of excitation voltage and unstrained voltage, 
error is quite sensitive to gage factor.
Figure 3-21. Error of Quarter Bridge Linear Approximation
Error vs Delta Strain
-1000
0
1000
2000
3000
4000
5000
6000
-100000
-80000
-60000
-40000
-20000
0
20000
40000
60000
80000
100000
Delta Strain in m icrostrain (µ ε)
Error in microstrain (µ
ε)
GF=1
GF=2
GF=5