Siemens syngo CT 2005C 사용자 설명서

다운로드
페이지 321
212
 
Children
ThoraxHRSeq_Child
Indications:
Sequence mode for High Resolution lung studies, e.g., 
interstitial changes of the lung parenchyma, etc.
A range of 20.1 cm is predefined.
* The conversion factor for a 7-year-old child, and a 
scan range of 201 mm was used.
ThorHRSeq
kV
130
Effective mAs
70
Rotation time
1.0 sec
Slice collimation
1.0 mm
Slice width
1.0 mm
Feed/Scan
10.0 mm
Kernel
U90s
CTDI
Vol
1.51 mGy
Effective dose
Male: 1.02 mSv*
 
Female: 1.21 mSv*
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book  Page 212  Friday, April 8, 2005  9:55 AM