Siemens syngo CT 2005C 사용자 설명서

다운로드
페이지 321
Shoulder
97
 
Hints
• Use raw data to review a target region if necessary.
• For image reconstruction of soft tissue, use kernel 
B31s and a slice width of 5.0 mm.
• Coronal and sagittal 2D planar reconstructions are 
important for evaluation of the joint space & bursa 
sacs in CT arthograms.
• 3D renderings are helpful for complex fractures & 
dislocations.
Shoulder
2
nd
 reconstruction
kV
130
Effective mAs
70
Rotation time
1.0 sec
Slice 
collimation
1.5 mm
Slice width
3.0 mm
2.0 mm
Feed/Rotation
3.0 mm
Pitch Factor
1.0
Increment
3.0 mm
1.5 mm
Kernel
B60s
CTDI
Vol
7.55 mGy
Effective dose
Male: 
0.94 mSv
 
Female: 
1.13 mSv