Teledyne API - T100 UV Fluorescence SO2 Analyzer  
SETUP Menu 
 
 
111 
5.9.2. ANALOG OUTPUT STEP TEST 
Analog Output is used as a step test to check the accuracy and proper operation of the 
analog outputs. The test forces all four analog output channels to produce signals 
ranging from 0% to 100% of the full scale range in 20% increments. This test is useful 
to verify the operation of the data logging/recording devices attached to the analyzer. 
Access the Analog Output Step Test from the DIAG Menu (refer to Figure 5-16), the
press:  
 
 
 
 
 
SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X    
 
 < TST  TST >  CAL  
 
 
SETUP
 
SAMPLE          
ENTER SETUP PASS : 818
 
 
        
 8        1        8 
 
                      
ENTR
  
 EXIT
 
SETUP X.X 
 
      
PRIMARY SETUP MENU
 
  
 
 CFG  DAS
  
RNGE
  
PASS  CLK  
MORE
  
   EXIT
 
 
 
 
DIAG  
SIGNAL I / O 
 
NEXT 
      
                                      
ENTR 
  
EXIT
 
Exit-Exit 
returns to the 
DIAG
 menu 
SETUP X.X     
SECONDARY SETUP MENU
  
 
 
 
COMM  VARS  
DIAG
   
      
                   EXIT
 
DIAG  
ANALOG OUTPUT 
 
PREV
 
NEXT
 
      
                     ENTR    
EXIT
 
DIAG AOUT
 ANALOG 
OUTPUT
 
   0% 
 
      
 
 
   
EXIT
 
Performs 
analog output 
step test. 
0% - 100% 
DIAG AOUT 
ANALOG OUTPUT
 
   [0%] 
 
      
 
 
  EXIT
 
Pressing the “0%” button while performing the test  will 
pause the test at that level. Brackets will appear around 
the value: example: [20%] Pressing the same key again 
will resume the test. 
 
Figure 5-18: 
DIAG – Analog Output Menu 
 
06807C DCN6650