Data Acquisition System (DAS) and APICOM 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
158 
The Channel Properties, Triggering Events and Data Parameters/Functions for these 
default channels are: 
 
NAME: PNUMTC 
EVENT: ATIMER 
PARAMETERS: 
STARTING DATE: 01-JAN-07 
SAMPLE PERIOD: 000:00:05 
REPORT PERIOD: 001:00:00 
NO. OF RECORDS: 360 
RS-232 REPORT: OFF 
COMPACT REPORT: OFF 
CHANNEL ENABLED: ON 
NAME: CALDAT 
EVENT: SLPCHG 
PARAMETERS: 
NO. OF RECORDS:200 
RS-232 REPORT: OFF 
COMPACT REPORT: OFF 
CHANNEL ENABLED: ON 
CAL HOLD OFF: OFF
PARAMETER: SMPFLW 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: SMPPRS 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: OFSET1 
MODE: INST 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: SLOPE1 
MODE: INST 
PRECISION:
STORE NUM SAMPLES OFF
 
PARAMETER: ZSCNC1 
MODE: INST 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: CONC1 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: STABIL 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
NAME: CONC 
EVENT: ATIMER 
PARAMETERS: 
STARTING DATE: 01-JAN-07 
SAMPLE PERIOD: 000:00:01 
REPORT PERIOD: 000:00:05 
NO. OF RECORDS: 4032 
RS-232 REPORT: ON 
COMPACT REPORT: OFF 
CHANNEL ENABLED: ON 
LIST OF CHANNELS
NAME: DETAIL 
EVENT: ATIMER 
PARAMETERS: 14 
STARTING DATE: 01-JAN-07 
SAMPLE PERIOD: 000:00:01 
REPORT PERIOD: 000:01:00 
NO. OF RECORDS:480 
RS-232 REPORT: OFF 
COMPACT REPORT: OFF 
CHANNEL ENABLED: ON 
NAME: FAST 
EVENT: ATIMER 
PARAMETERS: 14 
STARTING DATE: 01-JAN-07 
SAMPLE PERIOD: 000:00:01 
REPORT PERIOD: 000:00:01 
NO. OF RECORDS:360 
RS-232 REPORT: OFF 
COMPACT REPORT: OFF 
CHANNEL ENABLED: ON 
PARAMETER: UVDET 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: LAMPR 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: DRKPMT 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: DARKUV 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: CONC1 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: STABIL 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: STRLGT 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: RCTEMP 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: SMPPRS 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: BOXTEMP 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: HVPS 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: REFGND 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: REF4096 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: PMTDET 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
Same parameters &
settings as DETAIL
 
Figure 7-1: Default DAS Channels Setup 