Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Calibration Procedures 
 
 
203 
9.10. CALIBRATION OF OPTIONAL SENSORS 
This section presents calibration procedures for the O
2
 sensor option and for the CO
2
 
sensor option. 
9.10.1. O
2
 SENSOR CALIBRATION 
Calibration begins with connecting the zero and span gases, then setting the 
concentration values. 
9.10.1.1. O
2
 CALIBRATION SETUP 
Bottled gases are connected as follows:  
 
Figure 9-13: 
O
2
 Sensor Calibration Set Up 
O
2
 SENSOR ZERO GAS: Teledyne API recommends using pure N
2
 when calibrating 
the zero point of your O
2
 sensor option. 
O
2
  SENSOR SPAN GAS: Teledyne API recommends using 20.9% O
2
 in N
2
 when 
calibration the span point of your O
2
 sensor. 
 
9.10.1.2. SET O
2
 SPAN GAS CONCENTRATION
 
 
Set the expected O
2
 span gas concentration.   
This should be equal to the percent concentration of the O
2
 span gas of the selected 
reporting range (default factory setting = 20.9%; the approximate O
2
 content of ambient 
air).   