Teledyne Camera Accessories T100 사용자 설명서
Principles of Operation
Teledyne API - T100 UV Fluorescence SO2 Analyzer
300
13.5.2. SENSOR MODULE
Electronically, the T100 sensor module is a group of components that: create the UV
light that initiates the fluorescence reaction between SO
light that initiates the fluorescence reaction between SO
2
and O
3
; sense the intensity of
that fluorescence; generate various
electronic signals needed by the analyzer to
determine the SO
2
concentration of the sample gas (refer to Section 13.1.1), and sense
and control key environmental conditions such as the temperature of the sample gas and
the PMT.
the PMT.
Figure 13-15:
T100 Sensor Module
These components are divided into two significant subassemblies: the sample chamber
and the PMT assembly.
and the PMT assembly.
Figure 13-16 shows an exploded view of the sample chamber assembly
Figure 13-17shows an exploded view of the PMT Assembly