Teledyne Camera Accessories T100 사용자 설명서

다운로드
페이지 422
Principles of Operation 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
300 
13.5.2. SENSOR MODULE  
Electronically, the T100 sensor module is a group of components that: create the UV 
light that initiates the fluorescence reaction between SO
2
 and O
3
; sense the intensity of 
that fluorescence; generate various
 
electronic signals needed by the analyzer to 
determine the SO
2
 concentration of the sample gas (refer to Section 13.1.1), and sense 
and control key environmental conditions such as the temperature of the sample gas and 
the PMT.   
 
Figure 13-15: 
T100 Sensor Module 
These components are divided into two significant subassemblies: the sample chamber 
and the PMT assembly. 
Figure 13-16 shows an exploded view of the sample chamber assembly 
Figure 13-17shows an exploded view of the PMT Assembly 
06807C DCN6650