Teledyne API - T100 UV Fluorescence SO2 Analyzer  
SETUP Menu 
 
 
93 
5.4.3.1. SINGLE RANGE MODE (SNGL) 
The default range mode for the analyzer is single range, in which all analog 
concentration outputs are set to the same reporting range. This reporting range can be set 
to any value between 0.1 ppb and 20,000 ppb.  
While the two outputs always have the same reporting range, the span and scaling of 
their electronic signals may also be configured for different differently (e.g., A1 = 0-10 
V; A2 = 0-0.1 V). 
To select SNGLE range mode and to set the upper limit of the range, press: 
 
 
 
SAMPLE          
ENTER SETUP PASS : 818
 
 
        
 8        1        8 
 
                      
ENTR
  
 EXIT
 
 
 
 
SETUP X.X 
 
      
PRIMARY SETUP MENU
 
  
 
 CFG  DAS
  RNGE  
PASS  CLK  MORE  
EXIT
 
 
 
 
SETUP X.X 
RANGE CONTROL MENU 
 
 
 
 
MODE
 
 
 
 
 
 
SET  UNIT
 
 
 
 
EXIT
 
 
 
 
SETUP X.X 
RANGE MODE: 
SNGL
 
 
 
 
 
SNGL
 
 
 
 
 
 
IND    AUTO 
 
     ENTR  EXIT
 
EXIT
x 2 returns
to the main 
SAMPLE
 display 
 
 
 
SETUP X.X 
RANGE MODE: 
SNGL
 
 
 
 SNGL
 
 
 
 
IND    AUTO 
 
     ENTR
  EXIT
 
 
 
 
SETUP X.X 
RANGE CONTROL MENU 
 
 
 MODE
 
 
 
 
SET
  UNIT
 
 
 
 
EXIT
 
 
 
SETUP X.X
RANGE: 
500.0 Conc
 
 
 0          0          
5
         
0          0          .0
       
ENTR
  EXIT
 
 
 
 
SETUP X.X 
RANGE CONTROL MENU 
 
 
 MODE
 
 
 
 
SET  UNIT
 
 
 
 
EXIT
 
SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X  
 
 
 < TST  TST >  CAL  
 
 
SETUP
 
 
 
Figure 5-4: SETUP RNGE – Single Range Mode 
 
06807C DCN6650