SECTION 7—MAINTENANCE
29
Detecting Wear and Damage
It is important to inspect all stressed parts, such as slings, spreader bar and any pivot for 
slings for signs of cracking, fraying, deformation or deterioration. Replace any defective 
parts immediately and ensure that the lift is not used until repairs are made.
Cleaning the Sling and the Lift
The sling should be washed regularly in water temperature of 180°F (82°C) and a biological 
solution. A soft cloth, dampened with water and a small amount of mild detergent, is all that 
is needed to clean the patient lift. The lift can be cleaned with non‐abrasive cleaners.
Replacing the Electric Actuator
NOTE: For this procedure, refer to FIGURE 7.2.
1. Remove the nut, washer and shoulder bolt that secure the electric actuator to the mast 
mounting bracket.
2. Rest the boom on your shoulder and remove the nut, bolt, plastic bushing and pinch 
guard from the boom mounting bracket.
3. Remove the electric actuator assembly.
4. Reverse steps for installation.
CAUTION
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
Boom
Pinch Guard
Bushing
Mast