Getting Started with AMPS Test
Maximizing the Accuracy of Your Measurements
Maximizing the Accuracy of Your Measurements
This Test Set is designed to make highly accurate measurements. 
However, to ensure that you have the most accurate measurements 
available, you can perform the following tasks:
Calibration
You should calibrate the Test Set whenever you change or add an 
internal assembly. You may want to calibrate when you upgrade 
firmware. See the Assembly Level Repair Manual for calibration 
procedures.
TX Power Temperature Compensation
The Test Set is internally compensated. However, power measurement 
calibration can be optimized for temperature changes using the 
TX Pwr Zero
 field on the RF ANALYZER screen. The new calibration 
factors are stored in RAM until the next time the routine is used. 
To have the Test Set zero the TX Power measurement automatically 
when needed, set the 
Auto Zero
 field on the RF ANALYZER screen to 
Auto
. During operation, the Test Set will temporarily halt the TX 
Power measurement as it is calibrated. This can happen during a 
measurement. If interrupting the measurement is a problem for your 
test setup, set the 
Auto Zero
 field to 
Manual
, and select 
TX Pwr Zero
 whenever you want to manually calibrate the 
measurement.