2.3.8.1 Column tab 
 
The Column tab is for setting operating condition of the electron optical column. 
 
(1) HIGH MAG MODE block 
Set column condition for High Mag mode. 
 
(1-1) Probe current 
Select Normal or High. For resolution prior 
operation, use Normal. In case where larger probe 
current is required, select High. 
(1-2) Focus Mode 
Select UHR or HR. 
UHR mode is optimized for operation at short 
working distance. Full working distance range 
allowed for the present accelerating voltage can 
be used. 
HR mode is for use of rather long working 
distance. Longer than about 5mm working 
distance is able to be focused. 
(1-3) Cond Lens 1 
The check box On or Off excitation of the 1st 
Condenser Lens. 
Usually set it On (checked) 
Select a excitation value with selection box. The range is 1.0 to 16.0. Higher number 
results stronger excitation and then small probe current. Usual recommendation 
value is 5.0.   
For fine setting, use the up-down buttons. 
(1-4) Cond Lens 2 
The check box On or Off excitation of the 2nd Condenser Lens. 
Usually set it On (checked) 
(1-5) Focus Depth 
Set ocus depth. Selectable range is 0.5 to 2.0. Resolution optimized condition is 1.0. 
To increase focus depth, select larger number.   
For usual operation, set it to 1.0 or larger to increase focus depth. 
(1-6) Specimen Bias Voltage 
Specimen Bias Voltage (-15V) is applied for optimum secondary electron detection.   
Set it to ON for normal operation.    Turn it off for some special applications such as 
EBIC. 
 
When operating the S-4800 at short working distance, you may end up with images 
of uneven brightness throughout the field at low magnifications.    We suggest to set 
Specimen Bias Voltage at OFF in order to correct this problem. 
 
When the specimen is tilted, however, the upper detector signal will decrease