Hitachi Microscope & Magnifier S-4800 사용자 설명서

다운로드
페이지 294
 
3
-
104
(2) Operation Panel tab 
The arrangement of control blocks to be placed on the operation panel can be customized. 
   
 
(3) Tool bar Setting tab 
Select tool buttons to place on the tool button section of the 
operation panel. 
Tool buttons put check mark are shown on the tool button 
section. 
                                       
(4) Stage tab 
 
(a) MOTOR DRIVE STAGE 
Selects enable (On) or disable (Off) for motor driven 
stage (5 axes motorized stage of type II or 2 axes 
motor drive option of type I instrument). Set it at Off 
when the motorized stage is removed or not used, for 
example when using the cryo-stage. It is necessary 
because control program accesses motor drive part 
and may locked up if there is no response from the 
motorized stage. 
(b) REVERSE MODE 
Changes direction of image motion by mouse 
operation on the scanning image and by the trackball. 
When it is not checked, image moves to the direction 
of mouse movement or trackball rotation. When it is 
checked, image moves to reverse direction. 
(c) TRACK BALL / JOYSTICK SPEED 
When this box is checked, stage driving speed by the trackball or the joystick option can 
be adjusted with the Speed slider on the Stage tab on the operation panel. 
 
(5) MOUSE OPERATION tab 
Set the sensitivity of mouse operation (focus, stigma. 
brightness and contrast adjustment) on the scanning image.   
                                         
Set a slider control at the Slow side for lower sensitivity or 
at the Fast side for higher sensitivity. Adjustable range is a 
half to 1.5 times of default sensitivity. 
 
The IMAGE ADJUSTMENT area selects mouse operation 
functions to be enabled. Check to use the control