DIAGNOSTIC CODES
LOBAT
Weak batteries: change all batter-
ies as soon as possible.
-BA-
Exhausted batteries: change all bat-
teries immediately.
-LO-
Low light level is received during the
zeroing procedure: check the cuvet
for scratches and ensure that the
sample is not excessively turbid.
Repeat the reading. If the problem
persists, recalibrate the meter us-
ing deionized water in the diagnos-
tic mode “31” (see “Calibration”).
-CAP-
High light intensity during last mea-
surement: make sure that the cuvet
is capped and placed properly in
the holder, and that the ambient
light does not reach the photode-
tector. Repeat the measurement. If
the problem persists, contact your
dealer or the nearest HANNA office.
Er 1
Hardware error: repeat the measure-
ment. If the error message appears
again, contact your dealer or the
nearest HANNA office.
r n G
Out of range: check the measuring
procedure and verify the concentra-
tion of the sample to ensure that is
not too high.