4–20 mA Option
8990cm.fm
Section 6
5.
Select an analog Input Channel (e.g., channel 1, 2, 3, or, flow, etc.) to 
assign to that output. Press 
CHANGE CHOICE
 to cycle through the 
channel names. When the desired channel is displayed, press 
ACCEPT
.
6.
Assign a channel value to the 4 mA current value. This value is typically 0, 
however any value can be set. Enter the value of the input needed to 
generate 4 mA of current at the output.
7.
Assign an input value to the 20 mA current level.
8.
Repeat this process to configure the other 4–20 mA output.
6.3.3 4–20 mA Calibration
After wiring the 4–20 mA connection perform a 4–20 mA output calibration. 
The 4–20 mA output calibration requires a digital multimeter or access to the 
4–20 mA current loop wiring. Two 4–20 mA outputs are available and are 
designated Output A and Output B. Both outputs are calibrated the same way 
and are isolated from each other. 
Calibration may be performed while the 4–20 mA device is in the current loop, 
as shown in 
 or disconnected from the current loop as shown in 
. In either case, the multimeter must be set to a 20 milliamp dc range 
or greater.
1.
From the Main Menu, select 
OPTIONS > ADVANCED OPTIONS > 
CALIBRATION > 4–20 mA OUTPUTS.
2.
Connect a multimeter to the 4–20 mA current outputs per 
 an
3.
Make sure that the 4–20 mA output is enabled. If it is not enabled, press 
CHANGE CHOICE
 so that the display shows Enabled and press 
ACCEPT
.
4.
Select the output (A or B) to calibrate.
5.
Press any key to set the selected output to 4.00 mA dc.
6.
Measure the current on the selected output using the multimeter and 
enter the measured value using the numeric keypad. Press 
ACCEPT
.
11:00 AM 21 - APR - 01 
4–20 mA OUTPUTS
ACCEPT
CHANGE
CHOICE
INPUT CHANNEL:
CANCEL
FLOW
SELECT APPROPRIATE UNITS
11:00 AM 21 - APR - 01 
4–20 mA OUTPUTS
ACCEPT
4 mA INPUT VALUE
CLEAR
ENTRY
0.00 mgd
CANCEL
SELECT APPROPRIATE UNITS