2-18
Installation and Operation
5707070-A
Advanced Energy
®
To Operate Your Doppler Flow Meter for the First Time:
1. Complete all the steps in 
.
2. Complete all the steps in 
3. Turn on power and observe the display panel LCD. The display will show the 
model name and number of the unit.
4. Following the prompts presented by the display panel LCD, and making your 
entries using the keypad, configure your unit for operation.
4 The LCD screen displays SELECT RATE. Select the unit of measure for rate 
display:
If you select FPS, the flow meter will automatically begin measuring flow in feet 
per second. If you select keys 2 through 7, the LCD screen will ask that you enter 
your pipe ID in inches. See 
 to estimate your 
pipe ID.
After you enter your pipe ID, the LCD screen will then display SELECT 
TOTALIZER
. Select the unit of measure for totalizer display:
Attention
:
If your display flashes LOW BAT at any time during setup or 
programming, consult 
 for 
troubleshooting information. 
Note: If you are switching from battery (DC) power to AC power and the 
A/C Select switch selection does not match your incoming power level, 
you may see a LOW BAT warning in the front panel display. Adjust the 
A/C Select switch to the correct incoming power level (115 VAC or 
230 VAC), and this warning should disappear.
Table 2-4. Rate selection
Key
Selection
Definition
1
FPS
Feet per second
2
GPM
Gallons per minute
3
CFS
Cubic feet per second
4
MGD
Million gallons per day
5
MPS
Meters per second
6
LPS
Liters per second
7
CMS
Cubic meters per second
8
< = >
Use this key to move between rate screens