Sono-Trak™ Doppler Ultrasonic Flow Meter
Chapter
4
Specifications and Reference 
Information
The following sections provide specification and accessory information for the Sono-
Trak™ Doppler ultrasonic flow meter, as well as useful pipe and conversion 
information.
PERFORMANCE SPECIFICATIONS
Table 4-1. Performance specifications
Description
Specification
Fluid conditions
Measured fluid
Homogenous liquids with Newtonian flow profiles (sludges, 
slurries, sewage, wastewater, and other fluids containing 
suspended particles)
State of flow
Axis-symmetric flow in pipe filled with fluid
Fluid type
Liquids containing suspended particles of at least 35 ppm at 
40 microns at 0.2% density of undissolved solids or air bubbles
Measurable fluid 
temperature range
-40 ºF to 300 ºF (-40 ºC to 149 ºC)
Velocity range
0.1 fps to 50 fps (0.04 mps to 15.25 mps)
Flow sensitivity
0.01 fps (0.04 mps)
Piping conditions
Pipe diameter range
Velocity indication: 0.5" to 300" (13 mm to 7620 mm) diameter
Volumetric indication: 0.5" to 100" (13 mm to 2540 mm) 
diameter
Pipe
Including but not limited to: carbon steel, stainless steel, ductile 
iron, cast iron, FRP, PVC, fiberglass, teflon, and most concrete 
lined or coated pipes
Pipe lining
Including but not limited to: tar, epoxy, mortar, rubber, teflon, 
PVC, and glass