4-4
Specifications and Reference Information
5707070-A
Advanced Energy
®
Alarm output
When on AC power:
15.0 VDC = Active alarm
High alarm: switched +15.0 VDC, 500 mA
Low alarm: switched +15.0 VDC, 500 mA 
When on battery power:
12.0 VDC = Active alarm
High alarm: switched + 12.0 VDC, 500 mA
Low alarm: switched + 12.0 VDC, 500 mA
Power failure backup
An internal super-capacitor allows the electronics assembly to 
retain all parameter information and current totalizer value for 
up to 10 days
Standard interface
LCD display
2-line, 16 characters per line, high resolution
LED indicators
Yellow: Echo signal received
Red: High alarm or low alarm activated
Green: Power available
Keypad
12 keys, tactile feedback
Note: Keypad lockout available through internal switch; see 
 and 
Display panel
Rate selection, totalizer selection, K factor selection, analog 
output set, low flow cutoff set, update time (damping) set, high/
low alarm set, fps or mps selection, totalizer reset
Output setting function
4-20 mA output scaling (set in velocity from keypad)
Alarm setting function
High and low alarm set points (entered in velocity from keypad)
Rate indication
4-digit LCD, velocity or volume (user selectable)
English units: feet per second (fps), gallons per minute (gpm), 
cubic feet per second, and million gallons per day
Metric units: meters per second (mps), liters per second (lps), 
and cubic meters per second
Note: Gallons refers to U.S. gallons.
Table 4-3. Functional specifications (Continued)
Description
Specification