FMC Corporation - Talstar 80 사용자 설명서
Proline Promass 80 PROFIBUS PA
Troubleshooting
73
031
S: SENSOR HW DAT
$: # 031
$: # 031
0x10
0x11
0x12
0x11
0x12
BAD
Sensor Failure
O.K.
Low
High
S-DAT failure /
S-DAT not inserted
S-DAT not inserted
Cause:
1.
1.
S-DAT is not plugged into the amplifier board
correctly (or is missing).
correctly (or is missing).
2.
S-DAT is defective.
Remedy:
1.
Check whether the S-DAT is correctly plugged
into the amplifier board.
into the amplifier board.
2.
Replace the S-DAT if it is defective.
Check that the new, replacement DAT is
compatible with the measuring electronics.
Check the:
- Spare part set number
- Hardware revision code
compatible with the measuring electronics.
Check the:
- Spare part set number
- Hardware revision code
3.
Replace measuring electronics boards if
necessary.
necessary.
4.
Plug the S-DAT into the amplifier board.
032
S: SENSOR SW DAT
$: # 032
$: # 032
0x10
0x11
0x12
0x11
0x12
BAD
Sensor Failure
O.K.
Low
High
S-DAT data
inconsistent
inconsistent
121
S: A/C COMPATIB.
!: # 121
!: # 121
0x0F
BAD
Device Failure
Constant
Amplifier and I/O
board only partially
compatible
board only partially
compatible
Cause:
Due to different software versions, I/O board and
amplifier board are only partially compatible
(possibly restricted functionality).
Due to different software versions, I/O board and
amplifier board are only partially compatible
(possibly restricted functionality).
!
Note!
• This message is only listed in the error history.
• Nothing is displayed on the display.
• Nothing is displayed on the display.
Remedy:
Module with lower software version has either to
be actualized by FieldCare with the required
software version or the module has to be replaced.
Module with lower software version has either to
be actualized by FieldCare with the required
software version or the module has to be replaced.
261
S: COMMUNICAT. I/O
$: # 261
$: # 261
0x18
0x19
0x19
0x1A
BAD
No
Communicati
on
O.K.
Low
High
Communication
failure
failure
Cause:
Communication error. No data reception between
amplifier and I/O board or faulty internal data
transfer.
Communication error. No data reception between
amplifier and I/O board or faulty internal data
transfer.
Remedy:
Check whether the electronics boards are correctly
inserted in their holders
Check whether the electronics boards are correctly
inserted in their holders
379
S: LOW FREQ. LIM
$: # 379
$: # 379
0x0F
BAD
Device Failure
Constant
Frequence limit 1
Cause:
The measuring tube oscillation frequency is outside
the permitted range.
The measuring tube oscillation frequency is outside
the permitted range.
Causes:
Measuring tube damaged
Sensor defective or damaged
Measuring tube damaged
Sensor defective or damaged
Remedy:
Contact your Endress+Hauser service organization.
Contact your Endress+Hauser service organization.
No.
Device status message
(local display)
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
diagnostic
message in the
PROFIBUS
master
Cause/remedy (spare part
→ Page 80 ff.)
Q
u
a
lit
y co
de (H
EX
)
M
e
asured va
lue st
atus
Qua
li
ty st
a
tus
Qu
ality
substatus
Li