FMC Corporation - Talstar 80 사용자 설명서

다운로드
페이지 122
Proline Promass 80 PROFIBUS PA
Troubleshooting
73
031
S: SENSOR HW DAT
$: # 031
0x10
0x11
0x12
BAD
Sensor Failure
O.K.
Low
High
S-DAT failure / 
S-DAT not inserted
Cause:
1.
S-DAT is not plugged into the amplifier board 
correctly (or is missing).
2.
S-DAT is defective.
Remedy:
1.
Check whether the S-DAT is correctly plugged 
into the amplifier board.
2.
Replace the S-DAT if it is defective.
Check that the new, replacement DAT is 
compatible with the measuring electronics.
Check the:
- Spare part set number
- Hardware revision code
3.
Replace measuring electronics boards if 
necessary. 
4.
Plug the S-DAT into the amplifier board.
032
S: SENSOR SW DAT
$: # 032
0x10
0x11
0x12
BAD
Sensor Failure
O.K.
Low
High
S-DAT data 
inconsistent
121
S: A/C COMPATIB.
!: # 121
0x0F
BAD
Device Failure
Constant
Amplifier and I/O 
board only partially 
compatible
Cause:
Due to different software versions, I/O board and 
amplifier board are only partially compatible 
(possibly restricted functionality).
!
Note! 
• This message is only listed in the error history.
• Nothing is displayed on the display.
Remedy:
Module with lower software version has either to 
be actualized by FieldCare with the required 
software version or the module has to be replaced.
261
S: COMMUNICAT. I/O
$: # 261
0x18
0x19
0x1A
BAD
No 
Communicati
on
O.K.
Low
High
Communication 
failure
Cause:
Communication error. No data reception between 
amplifier and I/O board or faulty internal data 
transfer.
Remedy:
Check whether the electronics boards are correctly 
inserted in their holders 
379
S: LOW FREQ. LIM
$: # 379
0x0F
BAD
Device Failure
Constant
Frequence limit 1
Cause:
The measuring tube oscillation frequency is outside 
the permitted range.
Causes:
Measuring tube damaged
Sensor defective or damaged
Remedy:
Contact your Endress+Hauser service organization.
No.
Device status message
(local display)
PROFIBUS measured value status
Extended 
diagnostic 
message in the 
PROFIBUS 
master
Cause/remedy (spare part 
→ Page 80 ff.)
Q
u
a
lit
y co
de (H
EX
)
M
e
asured va
lue st
atus
Qua
li
ty st
a
tus
Qu
ality
 substatus
Li
m
its