Fisher & Paykel MR730 사용자 설명서

다운로드
페이지 64
49
Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001
B2
M
AINTENANCE 
S
CHEDULE 
C
HECKLIST
(PHOTOCOPY AND FILE IN MAINTENANCE RECORDS)
Type of service check:
q
 monthly
q
 3 monthly
q
 6 monthly
q
 yearly
Humidifier Serial Number:___________________________
Date: ______________
Step
Method
Pass
Fail
Monthly
Check Temperature Probe
Refer Appendix B1.1
q
q
Check Reusable Heater Wire
Refer Appendix B1.1
q
q
Check Heater Wire Connections
Refer Appendix B1.1
q
q
Check Power Cable
Refer Appendix B1.1
q
q
Perform Self-Test
Refer Appendix B1.1
q
q
Three Monthly
Perform Temperature Probe Test and Check
Temperature Probe
see § 4.3 and Appendix B
q
q
Annually
Performance Check
see § 6
q
q
Electrical Safety Checks - as required
Method
Pass
Fail
Earth resistance test
Measure the resistance from the earth pin on the mains plug to:
i The heaterplate.(0.2 
Ω
 maximum) (To get a good contact on the
anodised heaterplate, clip under the front edge of it.)
q
q
ii The earth stud on the base of the humidifier(0.2 
Ω
 maximum).
q
q
Insulation resistance test
Use a 500V DC tester to measure the resistance from the phase pin
on the mains plug to:
i The heaterplate (10 M
Ω
 minimum).
q
q
ii The earth stud on the base of the humidifier (10 M
Ω
 minimum).
q
q
Earth Leakage test
Has to be measured using an electrical safety tester as per IEC601.
Measure earth leakage with the earth probe of the tester on the
heaterplate. Earth leakage current is measured at normal operating
voltage and frequency of supply.
Should be 0.5 mA maximum in no fault condition.
q
q
Signature:______________________