3.5.4.  Fault Codes 
FAULT CODES ARE ADDITIVE.  For example: A fault code of <1800> indicates that both a Purge Flow 
Fault <1000> and a Zone Flow Fault <0800> have occurred. 
 
Code 
Name 
Description 
<0001> 
Box Temperature 
Fault 
Enclosure temperature is outside normal range (or IR detector has 
failed). Check that the monitor is not being subjected to extreme 
temperatures. Verify that the ventilation holes are not obstructed. 
Use the DIAGNOS  function  (page 26)  to check the Box 
Temperature. 
<0002> 
Bench 
Temperature Fault 
Optical bench is outside normal operating range (or IR detector has 
failed). Check that the monitor is not being subjected to extreme 
temperatures. 
<0004> 
Manifold Pressure 
Fault 
The manifold pressure is outside normal operating range (or IR 
detector has failed). Enter the DIAGNOS  function  (page 26)  and 
record ALL data. Call the factory with this information for further 
instructions. 
<0010> 
Loop Fault 
The 4–20 mA current loop is open, or there is a high resistance in the 
circuit. Check the wiring to the load/monitoring circuit. If this feature is not 
being used, a 100 ohm resistor must be connected to the 4–20 mA 
connector (page 16). 
<0100> 
Zero Filter Fault 
There is gas present in the purge air. Locate the monitor in an area 
containing fresh air, or install a purge air line. Refer to page 7. 
<0200> 
Gain Set Fault 
The digipot autotune sequence has failed. This fault will only occur 
on first boot up or after a firmware upgrade. Call the factory for 
further instructions. 
<0400> 
A/D Fault 
A fault has occurred in the analog-to-digital circuitry. Contact the 
factory with this information for further instructions. 
<0800> 
Zone Flow Fault 
Check for: 
A kink in the gas-sample line  or exhaust line; a blocked external 
termination filter; a blocked internal hydrophobic filter (page  11); a 
failed pump. 
<1000> 
Purge Flow Fault 
Check for: 
A kink in the purge line; a blocked external termination filter; a 
blocked internal hydrophobic filter (page 11); a failed pump. Once 
the blockage has been cleared, the monitor will return to normal 
operation after the monitor completes a purge cycle. 
<4000> 
Zero Fault 
The IR detector’s output voltage is out of tolerance. Enter the 
DIAGNOS  function  (page 26) and record ALL data. Call the factory 
with this information for further instructions. 
<8000> 
Clipping Fault 
The detector voltage may be out of tolerance. Use the DIAGNOS 
function (page 26) to check the IR detector voltage. Call the factory 
with this information for further instructions.