Siemens SC9000 사용자 설명서

다운로드
페이지 4
DU-DR-0016-SC9000 Rev A 6/4/02 GCX Corp.  Page 1 of 4 
GCX Mounting Assembly Operation/Installation Manual 
Dräger Model Cato
 ®
 with Siemens SC9000 Patient Monitor 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Patient Monitor – Left Side on Pivot Arm  /  Power Supply on Right Side 
 
The purpose of these instructions is to: 
 
 
1.  Describe attachment of GCX mounting hardware to the Cato anesthesia machine. 
 
2.  Describe attachment of the Siemens monitor to the GCX mounting hardware. 
 
WARNING:  USE OF MOUNTING HARDWARE AND MONITORING COMPONENTS OTHER THAN THOSE 
    DESCRIBED 
IN 
THIS 
DOCUMENT MAY RESULT IN SERIOUS INJURY DUE TO TIPPING OF THE  
    ANESTHESIA 
MACHINE. 
 
 
 
 
 
THIS CATO MOUNTING CONFIGURATION REQUIRES THE 40KG COUNTERWEIGHT WITH  
    EXTENDED 
WHEEL 
ADAPTER 
UPGRADE 
SET FROM DRÄGER (PART NUMBER:  86 018 68) 
 
Installer:  When installation is completed, provide these instructions to the end user for future reference. 
 
INSTALLATION 
 
Tools Required:  Philips head screw driver, Hex key (incl.). 
 
Mounting the Channel to Cato Edition 
 
Parts Required: 4" Channel, M6 x 16 mm FHMS screws (2). 
     Brass 
Plate 
with 
inserts 
from 
Dräger 
(M29110) 
 
1.  To attach the Channel to the Cato left side accessory track,  
 
locate the brass plate in the accessory track, align the  
 
mounting holes in the Channel with the threaded holes on  
 
the brass plate then run the M5 flat head screws (2) partly  
 
in. (Note:  Fixed Stop is at the bottom of the Channel) 
2.  Set the height of the Channel in the accessory track and  
 
tighten the M6 screws to hold in place. 
 
Prepare M Series Arm 
 
1)  Install the Tilt Limiting Stop screw. 
 
 
 
 
 
 
M6 x 16 mm (2) FHMS 
Accessory Track
Fixed Stop 
5/16-18 x ½” SHCS (1)