Philips HEARTSTART XLT M3500B 사용자 설명서
Specifications
13-3
13
Defibrillator (cont.)
AED Mode
AED Energy Protocol: Fixed Energy (150 Joules).
AED Shock Series: 2, 3, or 4.
Shock Series Timer: off, 30, 60, 90, 120, 150, 180, or 210 seconds.
Text and Voice Prompts: Extensive text audible messages guide user
through protocol.
through protocol.
Protocol: Follows pre-configured settings. Can be configured to meet
American Heart Association and European Resuscitation Council guide-
lines.
American Heart Association and European Resuscitation Council guide-
lines.
AED Controls: On, Off, Pause/Resume, Analyze/Stop Analysis, Shock,
Lead Select, SpO
Lead Select, SpO
2
, HR Alarms, Display Contrast, ECG Gain, Volume,
Print Strip, Print Summary, Annotated Mark Event, Manual Mode Entry.
Indicators: LCD Display for ECG waveform and text prompts, Audio
Alerts, Voice Prompts, QRS Beeper, Charging Tone, Charge Done Tone,
Printer.
Alerts, Voice Prompts, QRS Beeper, Charging Tone, Charge Done Tone,
Printer.
Armed Indicators: Charge Done Tone, Available Energy indicated on
display, Voice Message.
display, Voice Message.
Patient Analysis: Per protocol, evaluates patient ECG and signal quality
to determine if a shock is appropriate and evaluates connection impedance
for proper defibrillation pad contact.
to determine if a shock is appropriate and evaluates connection impedance
for proper defibrillation pad contact.
Shockable Rhythms: Ventricular fibrillation with amplitude > 100 uV
and wide complex ventricular tachycardia with rates greater than 150 bpm.
and wide complex ventricular tachycardia with rates greater than 150 bpm.
Sensitivity and Specificity: Meets AAMI guidelines.