CDMA Measurements - Screens and Control Fields
Ofs Trnsfer
Output Port
Ofs Trnsfer
This field transfers the time offset value from the time offset
measurement (displayed in
measurement (displayed in
TmOfs
) to the
Time Offset
field
automatically. Before selecting
Ofs Trnsfer
, you must make at least
one of the following code domain measurements:
• Power
• Timing
• Phase
Operating Considerations
When using this command programmatically, be sure a value for
Tm
Ofs
has been obtained from a power, timing, or phase measurement.
(One method of ensuring the measurement has been completed would
be to send an HP-IB command to query the results of the
be to send an HP-IB command to query the results of the
Tm Ofs
field
before sending the transfer command).
Screens on which this field is present
COD DOM (Code Domain Analyzer): fast power setup controls menu
Output Port
This field selects the output port for the signal from the Test Set’s RF
and CDMA generators.
and CDMA generators.
RF Out
selects the RF IN/OUT connector.
Dupl
selects the DUPLEX OUT connector.
Screens on which this field is present
CDMA GENERATOR
Main Menu