Chapter 5
Analog Measurements - Screens and Control Fields
RF Gen Freq
RF In/Ant
RF Gen Freq
This field sets the RF generator’s frequency.
This field is displayed on the screens listed below when the
RF
Display
field on the INSTRUMENT CONFIGURE screen is set to
Freq
.
Screens on which this field is displayed
SPEC ANL (When the Controls field is set to “RF Gen”)
RF GENERATOR
RF GENERATOR
RF In/Ant
This field selects the input port for the spectrum analyzer.
Operating Considerations
Maximum signal levels are printed on the connector panel.
CAUTION
Connecting a signal of >60 mW to the ANT IN port can
cause instrument damage (although internal protection
circuits can typically withstand a short-duration signal of
several Watts).
cause instrument damage (although internal protection
circuits can typically withstand a short-duration signal of
several Watts).
If the overpower circuit is triggered (signified by a
warning message at the top of the screen), remove the
signal from the ANT IN port, and reset using
warning message at the top of the screen), remove the
signal from the ANT IN port, and reset using
Meas Reset
(press and release the
Shift
key, then the
Hold
key) or turn
the Test Set off and on to reset it.
Using the ANT IN port with the
Sensitivity
field set to
High
can
result in uncalibrated operation (a message appears on the screen when
this happens). The purpose for the high sensitivity setting is to allow
you to look and listen to very low level signals when absolute accuracy
is not essential.
this happens). The purpose for the high sensitivity setting is to allow
you to look and listen to very low level signals when absolute accuracy
is not essential.
See the
for more
information.
Screens on which this field is displayed
SPEC ANL
Main Menu