Introduction and Specifications
Specifications
1
1-13
DC Current
Range
Resolution
Absolute Uncertainty
±
5°C from calibration temperature
For fields strengths 
>
1 V/m but 
<
3 V/m, add 1% of
range
Relative Uncertainty 
±
1°C
24 Hours
90 Days
180 Days
1 Year
24 Hours
90 Days
nA
±
 (ppm output + nA)
±
 (ppm output + nA)
220
 µ
A
2.2 mA
22 mA
.1
1
10
45 + 10
45 + 10
45 + 100
50 + 10
50 + 10
50 + 100
55 + 10
55 + 10
55 + 100
60 + 10
60 + 10
60 + 100
24 + 2
24 + 5
24 + 50
26 + 2
26 + 5
26 + 50
µ
A
±
 (ppm output +
 µ
A)
±
 (ppm output +
 µ
A)
220 mA
2.2A
[Note 1]
.1
1
55 + 1
75 + 30
60 + 1
80 + 30
65 + 1
90 + 30
70 + 1
95 + 30
26 + .3
40 + 7
30 + .3
45 + 7
5725A Amplifier:
11A
10
330 + 470
340 + 480
350 + 480
360 + 480
100 + 130
110 + 130
Secondary Performance Specifications and Operating Characteristics
Included in Uncertainty Specifications
Temperature Coefficient
 [Note 2]
Burden
Voltage
Maximum
Load
Noise
Range
Stability 
±
1°C 24
Hours
 10°-40°C
0°-10°C
 and
Compliance
Limits
Adder
 [Note 3]
For Full
Accuracy
Bandwidth
0.1-10 Hz
Bandwidth
10-10 kHz
40°-50°C
[Note 4]
pk-pk
RMS
V
±
 (ppm output + nA)
±
 (ppm output + nA)/°C
±
nA/V
ppm output + nA
nA
220
 µ
A
 2.2 mA
 22 mA
 220 mA
 2.2A
5 + 1
5 + 5
5 + 50
8 + 300
9 + 7
 µ
A
1 + .40
1 + 2
1 + 20
1 + 200
1 + 2.5
 µΑ
3 + 1
3 + 10
3 + 100
3 + 1
 µ
A
A3 + 10
 µ
A
10
10
10
10
3
[Note 5]
.2
.2
10
100
2
 µ
A
20k
2k
200
20
2
6 + .9
6 + 5
6 + 50
9 + 300
12 + 1.5
 µ
A
10
10
50
500
20
 µ
A
5725A:
±
 (ppm output +
 µ
A)
±
 (ppm output +
 µ
A)/°C
ppm output + 
µ
A
µ
A
11A
25 + 100
20 + 75
30 + 120
4
0
4
15 + 70
175
Notes:
Maximum output from 5700A terminals is 2.2A. Uncertainty specifications for 220
 µ
A and 2.2 mA ranges are increased by 1.3 X when
supplied through 5725A terminals.
Specifications are otherwise identical for all output locations.
1. Add to uncertainty specifications:
±
200 X I
2
 ppm for >100 mA on 220 mA range
±
10 X I
2
 ppm for >1A on 2.2A range
2. Temperature coefficient is an adder to uncertainty specifications. It does not apply unless operating more than 
±
5°C from calibration
temperature.
3. Burden voltage adder is an adder to uncertainty specifications that does not apply unless burden voltage is greater than 0.5V.
4. For higher loads, multiply uncertainty specification by:
1
+
0.1 x actual load
maximum load for full accuracy
5. 5700A compliance limit is 2V for outputs from 1A to 2.2A. 5725A Amplifier may be used in range-lock mode down to 0A.
Minimum output: 0 for all ranges, including 5725A.
Settling time to full accuracy: 1 second for
 µ
A and mA ranges; 3 seconds for 2.2A range; 6 seconds for 11A range; + 1 second for range
or polarity change
Overshoot: <5%