4.1  Channel
  
 
 
After entering the Configuration mode, select Channel and press the Enter key to get into 
Channel mode. Press AI,  DI, Math keys to select the Analog input, Digital input or 
Mathematics input. Press directional keys〈 〉at the bottom to select the channel. Afterwards, 
press directional keys ↑ ↓ ←   →   on the left hand side  to select the column.   After 
Configuration 4.1 to 4.6, press 
Back  key to return to real-time display, all configurations 
will be memorized. 
 
 
 
 
                                                
     Figure 4  –  2
  
 
Name:  It is to define the name for each channel.  The name is retricted by six characters. 
Press Enter, a keyboard and several keys appear. BackSP  key means backspace, Select 
key means to select a character or number, Caps on means characters in capital, and Caps 
off 
means characters not in capital.   
 
Desc The description about a specific channel on the display. 
 
Log Method: The method of logging measured data. Select the column, and then choose 
the Log method of Instant, Average, Minimum or Maximum data. 
 
Disable:   Select Disable while a specific channel is not required at this time. 
    Instant: logging in the last measured data at the sampling interval 
Average:  logging in averaged measured data at the sampling interval 
Minimum: logging in minimum measured data at the sampling interval  
    Maximum: logging in maximum measured data at the sampling interval