MR810 Technical Manual Revision C  REF 185042601
 
41 
 
D
D.3 E
LECTROMAGNETIC 
I
MMUNITY
 
Guidance and manufacturer’s declaration – electromagnetic immunity 
The MR810 is intended for use in the electromagnetic environment specified below.  The customer or the user 
of the MR810 should assure that it is used in such an environment. 
Immunity Test 
IEC60601  
test level 
Compliance Level 
Electromagnetic environment - 
guidance 
Electrostatic 
discharge (ESD) 
IEC 61000-4-2 
± 6 kV contact 
± 8 kV air 
± 6 kV contact 
± 8 kV air 
Floors should be wood, concrete or 
ceramic tile.  If floors are covered 
with synthetic material, the relative 
humidity should be at least 30 %. 
Electrical fast 
transient/burst 
IEC 61000-4-4 
± 2 kV for power 
supply lines 
± 1 kV for 
input/output lines 
± 2 kV for power 
supply lines 
± 1 kV for 
input/output lines 
Mains power quality should be that 
of a typical commercial or hospital 
environment. 
Surge 
IEC 61000-4-5 
± 1 kV differential 
mode 
± 2 kV common 
mode 
± 1 kV differential 
mode 
± 2 kV common 
mode 
Mains power quality should be that 
of a typical commercial or hospital 
environment. 
Voltage dips, short 
interruptions and 
voltage variations on 
power supply input 
lines 
IEC 61000-4-11 
< 5 % U
T
 
(> 95 % dip in U
T
 for 
0,5 cycle) 
 
40 % U
T
 
(60 % dip in U
T
 for 5 
cycles) 
 
70 % U
T
 
(30 % dip in U
T
 for 
25 cycles) 
 
< 5 % U
T
 
(> 95 % dip in U
T
 for 
5 seconds) 
< 5 % U
T
 
(> 95 % dip in U
T
 for 
0,5 cycle) 
 
40 % U
T
 
(60 % dip in U
T
 for 5 
cycles) 
 
70 % U
T
 
(30 % dip in U
T
 for 
25 cycles) 
 
< 5 % U
T
 
(> 95 % dip in U
T
 for 
5 seconds) 
Mains power quality should be that 
of a typical commercial or hospital 
environment.  If the user of the 
MR810 requires continued 
operation during power mains 
operation, it is recommended that 
the MR810 shall be powered from 
an uninterruptible power supply or 
battery. 
Power Frequency 
(50/60 Hz) magnetic 
field 
IEC 61000-4-8 
3 A/m 
3 A/m 
Power frequency magnetic fields 
should be at levels characteristic of 
a typical location in a typical 
commercial or hospital 
environment. 
Note: U
T
 is the a.c. mains voltage prior to application of the test level.