National Instruments 800 Series 사용자 설명서
Chapter 3
VXIpc 800 Series Configuration and Installation
© National Instruments Corporation
3-13
Ethernet Power-On Defaults
The VXIpc-800 Ethernet circuitry loads its power-on settings from an
EEPROM. You should not change switch S1 from its default setting;
however, the switch setting required to do so is included in Figure 3-14
for informational purposes.
EEPROM. You should not change switch S1 from its default setting;
however, the switch setting required to do so is included in Figure 3-14
for informational purposes.
Figure 3-14. Ethernet Power-On Defaults
Installing the VXIpc 800 Series
This section contains general installation instructions for the
VXIpc-800. Consult your VXIbus mainframe user manual or technical
reference manual for specific instructions and warnings.
VXIpc-800. Consult your VXIbus mainframe user manual or technical
reference manual for specific instructions and warnings.
1.
Plug in your mainframe before installing the VXIpc-800. The
power cord grounds the mainframe and protects it from electrical
damage while you are installing the module.
power cord grounds the mainframe and protects it from electrical
damage while you are installing the module.
Warning: To protect both yourself and the mainframe from electrical hazards,
the mainframe should remain off until you are finished installing the
VXIpc-800 module.
VXIpc-800 module.
2.
Remove or open any doors or covers blocking access to the
mainframe slots.
mainframe slots.
3.
If you are installing the VXIpc-800 into a D-size mainframe, install
a support designed for installing C-size boards in D-size
mainframes. The VXIpc-800 has no P3 connector and cannot
provide P3 Slot 0 control to VXI devices requiring this capability.
a support designed for installing C-size boards in D-size
mainframes. The VXIpc-800 has no P3 connector and cannot
provide P3 Slot 0 control to VXI devices requiring this capability.
Caution:
If the VXIpc-800 is not configured for automatic Slot 0 detection, be
certain that the slot you select in your VXIbus mainframe matches the
VXIpc-800 configuration as either a Slot 0 device or a Non-Slot 0 device.
If you install your VXIpc-800 into a slot that does not correspond with the
jumper setting, you risk damage to the VXIpc-800, the VXIbus backplane,
or both.
certain that the slot you select in your VXIbus mainframe matches the
VXIpc-800 configuration as either a Slot 0 device or a Non-Slot 0 device.
If you install your VXIpc-800 into a slot that does not correspond with the
jumper setting, you risk damage to the VXIpc-800, the VXIbus backplane,
or both.
S1
S1
a. Ethernet Power-On Defaults
Loaded from EEPROM
b. Hardwired Ethernet Values
!