Omega OM-USB-1208FS 사용자 설명서

다운로드
페이지 30
OM-USB-1208FS User's Guide 
Specifications 
23 
Note 1:  
Input current is a function of applied voltage on the analog input channels. For a given input 
voltage, V
in
, the input leakage is approximately equal to (8.181*V
in
-12) µA. 
Note 2:  
Maximum throughput scanning to PC memory is machine dependent. The rates specified are for 
Windows XP only. Maximum rates on operating systems that predate XP may be less and must 
be determined through testing on your machine 
Note 3:  
The AD7870 converter only returns 11-bits (0-2047 codes) in single-ended mode. 
Table 2. Accuracy, differential mode  
Range 
 Accuracy (LSB) 
±20 V  
5.1 
±10 V 
6.1 
±5 V 
8.1 
±4 V 
9.1 
±2.5 V 
12.1 
±2 V 
14.1 
±1.25 V 
20.1 
±1 V 
24.1 
Table 3. Accuracy, single-ended mode 
Range 
 Accuracy (LSB) 
±10 V 
4.0 
Table 4. Accuracy components, differential mode - All values are (±) 
Range 
% of Reading  
Gain Error at full scale (FS) 
(millivolts (mV)) 
Offset (mV) 
Accuracy at FS (mV) 
±20 V  
0.2 
40 
9.766 
49.766 
±10 V 
0.2 
20 
9.766 
29.766 
±5 V 
0.2 
10 
9.766 
19.766 
±4 V 
0.2 
9.766 
17.766 
±2.5 V 
0.2 
9.766 
14.766 
±2 V 
0.2 
9.766 
13.766 
±1.25 V 
0.2 
2.5 
9.766 
12.266 
±1 V 
0.2 
9.766 
11.766 
Table 5. Accuracy components, single-ended mode - All values are (±) 
Range 
% of Reading  
Gain Error at FS (mV) 
Offset (mV) 
Accuracy at FS (mV) 
±10 V  
0.2 
20 
19.531 
39.531 
Table 6. Noise performance, differential mode 
Range 
Typical counts  
Least significant bit
root mean square
 (LSB
rms)
 
±20 V 
0.30 
±10 V 
0.30 
±5 V 
0.45 
±4 V 
0.45 
±2.5 V 
0.61 
±2 V 
0.76 
±1.25 V 
1.06 
±1 V 