BiliTx Parents’ Manual
18
Light Source
Irradiance Level
Standard Panel-Light:   
 
30µW/cm
2
/nm
Neonatal Panel-Light: 
 
55µW/cm
2
/nm
Intensity Ratio
 
Standard Panel and Neonatal Panel : > .4 (minimum to maximum)
Audible Noise
 
< 60 dB(A). Measured in accordance with IEC 60601-2-50.
Standards Compliance
This device is designed to conform to the following standards: 
IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment
 
IEC 60601-2-50 Requirements for the Safety of Infant Phototherapy Equipment
 
Electromagnetic Compatibility: EN 60601-1-2, 2nd edition.
 
Electrical Requirements
AC Power
100-240 VAC, 50/60 Hz, 1.0 A
Type of Protection Against Electrical Shock
Class I Equipment
Degree of Protection Against Electrical Shock
 Type BF Applied Part
Degree of Protection Against Ingress of Water
Ordinary Protection, IPX0
Mode of Operation
Continuous
Disposal
Dispose of this device in accordance with local regulations.