HV-WP/HW-WP Series
14. Calibration (Adjusting Scale)
14.2.3.
14.2.3.
14.2.3.
14.2.3.  Calibration of the Zero Point
Calibration of the Zero Point
Calibration of the Zero Point
Calibration of the Zero Point
Step 8 Press and hold the 
 CAL 
 switch to enter
the calibration mode after displaying
normal weighing for 30 minutes.
Then 
 Cal 0 
 is displayed
Step 9 With nothing on the weighing pan, press
the 
 ENTER 
 switch while the stable mark is
displayed. The scale stores the current
condition as the zero point.
Step 10 The scale displays 
 5pn1 
 for several
seconds. Finishing the calibration mode
at this stage, proceed to step 14.
14.2.4.
14.2.4.
14.2.4.
14.2.4.  Span Calibration
Span Calibration
Span Calibration
Span Calibration
Step 11 Set the value of the calibration mass
using the following switches. (This initial
value is according to each product.)
 ∧ 
 switch
Selecting the number of a
figure.
 < 
 switch
Selecting a figure.
Step 12 Place the mass on the weighing pan
which was set at step 11, press the
ENTER 
 switch while the stable mark is
displayed. The scale then calculates the
span and stores it.
Step 13 The scale displays 
 end 
 at the finish.
Remove the mass from the weighing pan.
Step 14 Press the 
 CAL 
 switch to return to the normal
weighing mode.