A&D HW-60KV-WP 사용자 설명서

다운로드
페이지 72
Page 58
HV-WP/HW-WP Series
17.3. OP-04, Options
17.3.
17.3.
17.3.
17.3.  RS-422/ RS-485 / Relay output (OP-04)
RS-422/ RS-485 / Relay output (OP-04)
RS-422/ RS-485 / Relay output (OP-04)
RS-422/ RS-485 / Relay output (OP-04)
Replacing the RS-232C interface with this option, the RS-422/ RS-485 interface can
connect up to 16 scales and control them from a computer.
The RS-422/ RS-485 interface has the following two modes.
Steam mode
Outputs data countinuouslly.
Command mode
Controls the scale using commad from a computer.
Set the parameters for the "Baud rate ( 
 f4 
 )", " Output mode ( 
 f5 
 )" and " Format
 f15 
 )", in the function table, in advance.
Transmission system
EIA RS-422/ RS-485
Transmission
Asynchronous, bi-directional, half-duplex
Data format
Baud rate:
2400, 4800, 9600 bps
Data:
7 bits
Parity:
1 bit, Even
Start bit
1 bit
Stop bit
1 bit
Code
ASCII
Terminator
CR LF   (CR: 0Dh, LF: 0Ah)
Address
01 ~ 99
Address parameter ( 
 f11 
 ) of the function table.
Relay
Solid-state-relay
Maximum voltage
DC50V
Maximum current
DC100mA
 Maximum 
resistance
Ω
Adaptable connector
TM:BLA9 (of accessory)
Pin connections
Pin No. Signal name
Direction
Description
A
1
SDA
Output
RS-422/485 transmission A terminal
A
2
SDB
Output
RS-422/485 transmission B terminal
A
3
RDA
Input
RS-422/485 receive A terminal
A
4
RDB
Input
RS-422/485 receive B terminal
A
5
TRM
-
120
Ω
 terminator
A
6
HI
Output
Relay output of HI
A
7
OK
Output
Relay output of OK
A
8
LO
Output
Relay output of LO
A
9
COM
-
Relay common terminal
A