Stimulus registration method
i. Select the type of the start point to register: Start 1 or Start 2. (Start 2 can be selected only if “Blind spot” is set in
Start 2.)
ii. Select the intermediate filter (a, b, c, d, or e), intensity (1, 2, 3, or 4), and size (I, II, IV, or V) of the stimulus to be
registered.
iii. Press [Register], and the stimulus is registered at the selected start point.
Example) To register “I/3e” for Start 2
i. Select Start 2.
ii. Select “e” for the intermediate file, “3”
for the intensity, and “I” for the size.
iii. Press [Register].
iv. “I/3e” is added to the registered
stimulus.
Stimulus deletion method
i. Select the type of the start point to delete: Start 1 or Start 2. (Start 2 can be selected only if “Blind spot” is set in
Start 2.)
ii. Select the stimulus to delete from the “Stimulus” area.
iii. Press [Delete], and the selected stimulus is deleted.
Example) To delete “I/2e” for Start 1 in the above settings
i. Select Start 1.
ii. Select “I/2e”
iii. Press [Delete], and the selected
stimulus is deleted.
iv. “I/2e” is deleted from the registered
stimulus.
4
Other
When there is no response from the patient up to the end of the examination, “×” is displayed.
When “Test” is set to “Auto” and there are multiple isopters, the examination is performed beginning with the
stimulus which is easily visible to the patient.
stimulus which is easily visible to the patient.
When “Test” is set to “Auto” and there are multiple isopters, the examination start position is determined
based on the examination result if the isopter is the second or subsequent one and there is an examination
result in the same examination direction.
based on the examination result if the isopter is the second or subsequent one and there is an examination
result in the same examination direction.