Disassembly Procedures
7-10 
Disassembly  
Rear Case 
Disassembly
You must disassemble the rear case to remove the following modules:
Main  PCB
Patient Monitoring I/O
SpO
2
CO
2
NBP
Communications
Power Supply
Caution
Before you remove any module, make sure you are wearing ESD protection and you are 
working in a grounded environment.
To disassembly the rear case:
Step
Action
1
Verify that the battery has been removed. If not see, “Removing the Battery on 
page 7-4
.
2
Verify that the recorder has been removed. If not see, “Removing the Optional, 
External Recorder on page 7-6
. If there is no recorder, unsnap the recorder door 
using a flathead screwdriver (see figure below).