Philips 862474 Manual Do Utilizador

Página de 182
General Site Preparation Safety Information
2-2 
Site Preparations  
General Site Preparation Safety Information
Warning
Warning
To avoid contaminating or infecting personnel, the service environment or other 
equipment, make sure that equipment which has been used before has 
been appropriately disinfected and decontaminated.
Disconnect the monitor from the AC source by unplugging the power cable from the AC 
power connector located on the rear of the monitor. The On/Standby 
button does not disconnect the monitor from the AC mains supply
Accessory equipment connected to the monitor’s data interface must be certified 
according to IEC Standard 60950 for data-processing equipment or 
IEC Standard 60601-1 for electromedical equipment. All combinations 
of equipment must be in compliance with IEC Standard 60601-1-1 
systems requirements.
Explosion Hazard. Equipment not suitable for use in the presence of a flammable 
anaesthetic mixture with air, or with oxygen or nitrous oxide.
Make sure that you have read all applicable instructions before attempting to install the 
wall mount.
Wall mounts that are intended to support monitors must be capable of supporting 4 
times the weight of the monitor when properly installed. 
Do not mount any portion of a monitoring instrument over a patient’s bed.
Do not exceed the maximum rated load specified for each wall mount.
Ensure that no electrical wiring, utilities, or piping interfere with the selected wall 
mounting location.
Note
Do not return sensors, patient cables, NBP tubing and cuff, or the power cord.