Performance Procedures
4-30 
Testing the Monitor  
Documenting
 
CO
2
 Test 
Results
The following table lists the CO
tests and summarizes how to document the test results.
Test
Expected Test Results
What to record on service record
Barometric Pressure Check
X1 = difference between the reference 
pressure and the measured ambient 
pressure displayed on the monitor
(-12 mmHg < X1 < +12 mmHg)
PCO
2
: P/X1/X2/X3/X4/X5/X6/X7
or
PCO
2
: F/X1/X2/X3/X4/X5/X6/X7
P = passed, F = failed
X1: xx (two digits)
X2: x.x
X3: x.x
X4: xxx
X5: x.x
X6: x.x
X7: x.x
Leakage Check parts 1 and 2
X2 = value of part 1 leakage check on 
flowmeter (X2 </= 4.0 ml/min)
X3 = value of part 2 leakage check on 
flowmeter (X3</= 4.0 ml/min)
Pump Check
X4 > 120 mmHg below the ambient 
pressure
Flow Rate Check
X5 = CO
2
 flow rate (42.5 ml/min <X5 
<65 ml/min).
CO
2
 Gas Measurement Calibration 
Check
X6 = difference between measured 
CO
2
 value and calculated value, based 
on 5% CO
2
 cal. gas (X7<2.6 mmHg, 
or 0.35 kPa)
Calibration Verification
X7 = difference between measured 
CO
2
 value and calculated value, based 
on 10% CO
2
 cal. gas.
(X7±{0.07 x value calculated})