Micro Motion
®
 Series 3000 MVD Transmitters
Configuring the NOC Application
Figure 4-1
Well Performance Measurement menu
4.3
Basic configuration procedure
To configure the NOC application, the following general steps are required:
1. Set 
Mode of Operation
 to 
Continuous
 or 
Well Test
.
In Continuous mode, one well, separator, or pipeline is monitored continuously. A reset 
function is used to define the starting point for totals, averages, and maximum/minimum 
values.
In Well Test mode, a well test may be performed on any of the wells on a manifold, up to 
48 wells. Each well is configured independently and well test data is stored separately.
2. Set 
Reference Temperature
 to the reference temperature to be used by the NOC application.
3. If you are using Continuous mode, use the Well Data-Densities menu to set the well 
Continuous Mode
Well Test Mode
Well Performance Meas
60 degF
15 degC
20 degC
Transient Bubble Remd
• Drive  Gain  Level
• Action Taken
• Time Period
Transient Mist Remed
(2)
• Drive  Gain  Level
• Time Period
• Add Carry-Over Totals
• Gas Meter HART Tag
Press Comp Oil Dens
Shrinkage Factors
• Factor  for  Oil
• Factor for Water
Mode of operation
Reference Temperature
Well Data-Densities
Compensations
Wells 1 to 12
(1)
Wells 13 to 24
(1)
Wells 25 to 36
(1)
Wells 37 to 48
(1)
Well Name
(1)
Oil Density
Water Density
Gas Reference Density
Press Comp Oil Density
Reference Pressure
Purge Time
(1)
Oil Deviation
Water Deviation
Oil Duration Ave
Water Duration Ave
Well #: Well Name
(1)
Well #: Well Name
(1)
.
.
.
Well #: Well Name
(1)
Gas Meter
Gas Volume Units
Water Cut Monitor
Ext Water Cut Limit
External inputs
(1) Displayed only in Well Test mode.
(2) Displayed only if the Transient Mist Remediation option was purchased.